Seamaty Hematology
Part Auto Hematology Analyzer
Compact & Fully-Automated Bench Top Analyzer
- Efficient testing with precision reference laboratory quality results
- Compatibility with two modes of sampling and various printing options
- User-friendly with auto-cleaning of sample probe and tubes and automatic fault processing function
- Low consumption of only three reagents and ≤20uL of sample
- Lightweight design for easy portability
– 2 Reagents
– 21 Parameters
– 60 Samples/hour
– Low consumption of reagent and sample
Product Feature
Compatibility
Two modes of sampling: Whole blood, pre-diluted blood. Build-in thermal printer, external laser printer or inkjet printer.
User-friendly
10.4 colorful touchscreen. Liquid Crystal Display. Resolution : 800600. 10°incline adapt to visual observation
Less Maintenance
Auto-cleaning of sample probe and tubes. Automatic fault processing function
Low Consumption
Only need three reagents. Sample consumed ≤20uL
Efficient
60 samples per hour
Lightweight
390mm(L)×285mm(W)×390mm(H) Weight≤16Kg
Product Video
Specifications
- WBC
- RBC
- HGB
- MCV
- PLT
● Precision – ≤2.0% (3.5-15.0×10^9)/L
● Linear Range – (0-100.0×10^9)/L
● Contamination Rate – ≤0.5%
● Precision – ≤1.5% (3.0-6.0×10^12)/L
● Linear Range – (0.10-8.00×10^12)/L
● Contamination Rate – ≤0.5%
● Precision – ≤1.5% (100-180) g/L
● Linear Range – (0-250) g/L
● Contamination Rate – ≤0.6%
● Precision – ≤1.0% (70-120)fL
● Linear Range – ( – )
● Contamination Rate – ( – )
● Precision – ≤4.0% (150-500×10^9)/L
● Linear Range – (0-1000×10^9)/L
● Contamination Rate – ≤1.0%
Power
Voltage: AC 100V~240V, 50/60Hz; Rated Power: 120VA
Principle
Electric impedance method and Colorimetry method
Parameters
WBC, LYM#, MID#, GRAN#, LYM%, MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC 3 histograms
Reagents
Build-in LH Lyse, External Diluent, Probe Cleaner for maintenance
Flag System
RBC, WBC, PLT abnormal sample and support customized test groups
Storage
Up to 50,000 results including histograms, scattergrams and patient information.
Quality Control
3 Level QC, L-J graph, X-B
Calibration
Manual and Auto-calibration
Accessories
Standard: keyboard, waste liquid barrel, ground wire(3m); Option: mouse, external barcode scanner
Operational Environment
Working Temperature: 10℃~35℃; Relative Humidity: 20%~85%